Features:
Performance
- Microprocessor controller and temperature calibration function allows for high precision temperature control.
– Sterilization temperature at 110 to 123°C
– Melting* temperature at 60 to 100°C
* Liquefaction of coagulated agar media. - Automatic sterilization system for unattended operation.
– Heating > Check safe > Sterilization > Sterilization safe > Exhausting > End - Convenient operation modes for various uses.
– Standard modes and program modes are listed in detail as below tables. - Temperature calibration function.
Standard modes | Temp. (°C) | Time (min.) |
---|---|---|
Decontamination | 121 | 35 |
Solids sterilization | 121 | 20 |
Liquids sterilization | 121 | 15 |
Melting | 100 | 30 |
Program modes | Temp. (°C) | Time (min.) |
---|---|---|
Sterilization 1 | 110 to 123 | 1 to 999 |
Sterilization 2 | 110 to 123 | 1 to 999 |
Liquid sterilization | 110 to 123 | 1 to 999 |
Melting | 60 to 100 | 1 to 999 |
Convenience
- Conveniently monitored by an independent control display, sterilization process display, and pressure gauge.
– VFD for visualization of temp, pressure (optional), timer, warning alert checks.
– LED lamps for sterilization process check.
– Easy pressure check with the standard gauge attached to the front. - Vertical type autoclaves which allow for top loading are suitable for high volume sterilization.
– Baskets are stackable two high or even more levels, which secure inside the chamber space requirements. - Low-maintenance
– The chamber is constructed of stainless steel which is superiorly resistant to corrosion.
– Easy to replace silicone gasket. - Convenient documentation with the optional data recorder.
– Dot type recorder(6 point) allows for precise documentation.
Safety
- Mechanical safety devices
– Over current protection.
– Over temperature protection by automatic power cutoff.
– Over pressure protection by automatic and manual safety valve. - Safety warning systems
– Over temperature warning.
– Low temperature warning.
– Low water level warning.
– Low heat warning if the temperature does not reach the sterilization or melting temperature.
– Sterilization fail warning if the operation is unexpectedly stopped during sterilization process. - Slide opening door with plastic cover of low thermal conductivity prevents burns from the contact.
Specification:
-
Model ST-50G ST-65G ST-85G ST-105G Control Microprocessor PID control Chamber volume (L / cu ft) 50 / 1.8 65 / 2.3 85 / 3.0 105 / 3.7 Temperature 1) Sterilization temp. (℃ / ℉) 110 to 123 / 198 to 221.4 Melting temp. (℃ / ℉) 60 to 100 / 108 to 180 Pressure Sterilization at 121℃ (kg/cm2) 1.1 to 1.2 Gauge range in analog (Mpa) – 0.1 to 0.6 Safety Over temperature protector / Pressure safety valve / Over current protector Warning alert (over temp. / low temp. / low heat / sterilization fail / low water level) Timer 0 to 999 min. Dimension No. of baskets (standard / max.) 2 / 2 2 / 2 2 / 2 2 / 2 Load per baskets (Kg / lbs) 10 / 22.0 Basket (ØxH) (mm / inch) 380×180 / 14.9×7.1 380×230 / 14.9×9.1 380×330 / 14.9×13.0 380×405 / 14.9×15.9 Interior (WxH) (mm / inch) 400×458 / 15.7×18.0 400×558 / 15.7×21.9 400×758 / 15.7×29.8 400×908 / 15.7×35.7 Exterior (W×D×H) (mm / inch) 624x672x870 / 24.6×26.5×34.2
624x672x971 / 24.6×26.5×38.2
624x672x1083 / 24.6×26.5×42.6
624x672x1233 / 24.6×26.5×48.5
Exterior with exhaust tank (W×D×H) (mm / inch)
722x672x870 / 28.4×26.5×34.2
722x672x971 / 28.4×26.5×38.2
722x672x1083 / 28.4×26.5×42.6
722x672x1233 / 28.4×26.5×18.5
Net weight (Kg / lbs) 94.5 / 208.3 99.5 / 219.4 104 / 229.3 110 / 242.5 Electrical requirements (230V, 50 / 60 Hz) 14.3A 14.3A 14.3A 21.3A Cat. No. AAHL1015K AAHL1025K AAHL1035K AAHL1045K 1) Technical data according to EN 554 / BS EN 285:1997.