Features:

Performance

  • Microprocessor controller and temperature calibration function allows for high precision temperature control.
    – Sterilization temperature at 110 to 123°C
    – Melting* temperature at 60 to 100°C
    * Liquefaction of coagulated agar media.
  • Automatic sterilization system for unattended operation.
    – Heating > Check safe > Sterilization > Sterilization safe > Exhausting > End
  • Convenient operation modes for various uses.
    – Standard modes and program modes are listed in detail as below tables.
  • Temperature calibration function.
Standard modes Temp. (°C) Time (min.)
Decontamination 121 35
Solids sterilization 121 20
Liquids sterilization 121 15
Melting 100 30
Program modes Temp. (°C) Time (min.)
Sterilization 1 110 to 123 1 to 999
Sterilization 2 110 to 123 1 to 999
Liquid sterilization 110 to 123 1 to 999
Melting 60 to 100 1 to 999

Convenience

  • Conveniently monitored by an independent control display, sterilization process display, and pressure gauge.
    – VFD for visualization of temp, pressure (optional), timer, warning alert checks.
    – LED lamps for sterilization process check.
    – Easy pressure check with the standard gauge attached to the front.
  • Vertical type autoclaves which allow for top loading are suitable for high volume sterilization.
    – Baskets are stackable two high or even more levels, which secure inside the chamber space requirements.
  • Low-maintenance
    – The chamber is constructed of stainless steel which is superiorly resistant to corrosion.
    – Easy to replace silicone gasket.
  • Convenient documentation with the optional data recorder.
    – Dot type recorder(6 point) allows for precise documentation.

Safety

  • Mechanical safety devices
    – Over current protection.
    – Over temperature protection by automatic power cutoff.
    – Over pressure protection by automatic and manual safety valve.
  • Safety warning systems
    – Over temperature warning.
    – Low temperature warning.
    – Low water level warning.
    – Low heat warning if the temperature does not reach the sterilization or melting temperature.
    – Sterilization fail warning if the operation is unexpectedly stopped during sterilization process.
  • Slide opening door with plastic cover of low thermal conductivity prevents burns from the contact.

Specification:

  • Model ST-50G ST-65G ST-85G ST-105G
    Control Microprocessor PID control
    Chamber volume (L / cu ft) 50 / 1.8 65 / 2.3 85 / 3.0 105 / 3.7
    Temperature 1) Sterilization temp. (℃ / ℉) 110 to 123 / 198 to 221.4
    Melting temp. (℃ / ℉) 60 to 100 / 108 to 180
    Pressure Sterilization at 121℃ (kg/cm2) 1.1 to 1.2
    Gauge range in analog (Mpa) – 0.1 to 0.6
    Safety Over temperature protector / Pressure safety valve / Over current protector
    Warning alert (over temp. / low temp. / low heat / sterilization fail /  low water level)
    Timer 0 to 999 min.
    Dimension No. of baskets (standard / max.) 2 / 2 2 / 2 2 / 2  2 / 2
    Load per baskets (Kg / lbs) 10 / 22.0
    Basket (ØxH) (mm / inch) 380×180 / 14.9×7.1 380×230 / 14.9×9.1 380×330 / 14.9×13.0  380×405 / 14.9×15.9
    Interior (WxH) (mm / inch) 400×458 / 15.7×18.0 400×558 / 15.7×21.9 400×758 / 15.7×29.8  400×908 / 15.7×35.7
    Exterior (W×D×H) (mm / inch) 624x672x870

    / 24.6×26.5×34.2

    624x672x971

    / 24.6×26.5×38.2

    624x672x1083

    / 24.6×26.5×42.6

     624x672x1233

    / 24.6×26.5×48.5

    Exterior with exhaust tank

    (W×D×H) (mm / inch)

    722x672x870

    / 28.4×26.5×34.2

    722x672x971

    / 28.4×26.5×38.2

    722x672x1083

    / 28.4×26.5×42.6

     722x672x1233

    / 28.4×26.5×18.5

    Net weight (Kg / lbs) 94.5 / 208.3 99.5 / 219.4 104 / 229.3  110 / 242.5
    Electrical requirements (230V, 50 / 60 Hz) 14.3A 14.3A 14.3A  21.3A
    Cat. No. AAHL1015K AAHL1025K AAHL1035K  AAHL1045K

    1) Technical data according to EN 554 / BS EN 285:1997.
    * Permissible environmental conditions: temperature (5-40°C) and relative humidity (10-80%).